Specifications
- Detector Technology:
Photoionization, with 10.6 eV RF-excited electrodeless discharge tube
- Frequency Response: 330 Hz (@ the
-3 dB point)
- Detection Limit: 50 ppb
(propylene)
- Operating Concentration Ranges: Low
Gain: 0 - 1000 ppm, Med Gain: 0 - 200 ppm, High Gain: 0 - 40 ppm
- Precision: 5.0% (all gain
ranges)
- Sampling Rate: 1.1
litres/minute
- Operating Humidity Range: 0 to 100%
RH (non-condensing)
- Operating Temperature Range: 32F to
105F (0C to 40C)
Dimensions
- Sensor Head: 3.0" (7.6 cm) long,
2.0" (5.1 cm) wide, 1.0" (2.5 cm) thick.
- Controller: 5.25" (13.3 cm) long,
5.25" (13.3 cm) wide, 2" (5.1 cm) high.
Weight
- Sensor Head: 0.38 lbs (0.17 kg).
- Controller: 1.47 lbs (0.67 kg).
Power: 110/220VAC,
50/60 Hz.
Drawings
Drawings of the miniPID Fast-Response
Gas Sensor can be downloaded in Portable Document Format (PDF) and viewed using
Adobe Acrobat Reader. Acrobat Reader can be downloaded for free from Adobe.
Click here to download
Acrobat Reader.
Click on the drawing you would like to
download.
miniPID Outline Drawing - Outline drawing of
miniPID. miniPID Controller - Front Panel
View - Outline drawing of the Front Panel of the miniPID Controller.
miniPID Controller - Back Panel View -
Outline drawing of the Back Panel of the miniPID Controller.
References
The following are open-literature
papers describing data and data analysis obtained using miniPID or digitalPID
detectors in full-scale field trials.
miniPID and
digitalPID References.
|