The 322C High Speed Length Controller measures and controls the length of all types of muscle and connective tissue. This instrument is ideally suited, when matched with a 400A series Force Transducer, to measure length-tension, force-velocity, force-frequency, stiffness, work, kTR, force-pCa in muscle tissue.
The 322C provides up to 100 mN (10 grams) force and 6 mm of movement which makes it ideal for studying muscle fibers, bundles, strips and small whole muscle. The instrument is suitable for measuring skeletal, cardiac and smooth muscle.

Length Specifications
- Torque Arm Length: 2.0 cm
- Length Excursion: 6 millimeters
- Length Signal Resolution: 0.5 micron
- Length Signal Linearity: 0.1% over the center 1.2 millimeters , 0.5% over the entire 6 mm range
- Length In Scale Factor: 0.30 millimeters per volt ±10%
- Length Out Scale Factor: 0.30 millimeters per volt ±10%
- Length Step Response Time (1% to 99% critically damped) : 425.0 microseconds
- Dither at Lever Tip: 0.2 microns peak-to-peak maximum
- Sinusoidal Frequency Response (-3dB): 1.5kHz
Force Specifications
- Maximum Isometric Force: 100.0 mN (10 grams)
General Specifications
- Warm Up Time: 1 minute to rated accuracy
- Power Requirements: 120VAC ±10%, 50/60Hz, 35 Watts max. 100, 220 and 240VAC available.
- Dimensions Electronics Enclosure: 21.2cm wide x 25.3cm deep x 8.8cm high (8.375" x 10" x 3.5")
- Weight: Electronics: 2 kg, Motor: 42 grams
- Motor Cable Length: 10 feet.
- Power Cable Length: 6 feet.