Manuals - Bulletins - Drawings
Features
- Constant Current / Constant Voltage Operation
- 0.4 mA to 1 Amp @ up to ±80 Volts
- Positive, Negative and Bi-Phase Output Pulses
- Two Trigger Modes: Follow & Manual Trigger
- DMC, and ASI600A software include controls for the 701C
- Pulse Phase, Current/Voltage mode, Current/Voltage amplitude adjustable with front panel knobs.
- On/Off Safety Switch for Output Pulses
- Manual Trigger provides twitch pulse of fixed 1.25 millisecond duration
- 3 Signal Monitors provided: Sync Output, Current Output and Voltage Output
- Included with ASI muscle test systems: 1200A, 1300A, 1500A
The model 701C is a high-power, bi-phase stimulator that can provide positive, negative or alternating (bi-phase) pulses of up to 80V at up to 1 ampere of current. The high power output makes this stimulator ideal for field stimulation of muscle in a bath.
The stimulator can output either constant current pulses (from 0.4 milliamp to 1 amp) or constant voltage pulses (from 0 to 80 volts). The output current or voltage is set using a range selector switch and a graduated potentiometer, both located on the front panel.
The stimulator includes three output monitors to allow the researcher to monitor the timing (Sync Monitor) the voltage (V Monitor) or the current (I Monitor).
The stimulator provides two trigger methods:
- Pulse Trigger – the stimulator produces output pulses that match the timing and duration of the Pulse Trigger Input. This capability of "following" the input allows the stimulator to easily be controlled by a computer. The width and frequency of the stimulator's output pulses follow the width and frequency of the Pulse Trigger Input.
- Manual Trigger – A single pulse of 1.25 milliseconds duration is produced when the Manual Trigger Switch is pressed or held. Either a momentary push or a sustained hold of the trigger button will issue just a single pulse.
Aurora Scientific Inc. has two software products available (600A Digital Controller and Dynamic Muscle Control (DMC) software) that provide stimulator output control via simple on-screen controls. Precise timing and synchronization of stimulation and muscle lever control is assured when you combine our software control programs with a 701C stimulator. Each program provides complete control of all timing parameters affecting muscle stimulation including: pulse width, pulse frequency (pulse trains), train frequency and train duration.
A unique feature of the 701C stimulator is the Pulse Phase Switch that is used to set the polarity of the output pulses. The stimulator can be set to generate only positive pulses, only negative pulses, or pulses of alternating (positive then negative) bi-phase polarity. These pulses are all referenced to the chassis ground. Operation in the Bi-Phase mode allows the net current through the muscle bath to be zero. This produces better muscle stimulation as it minimizes corrosion of the stimulation electrodes and the potential release of heavy metals into the bath.
The 701C stimulator is housed in a half rack-mount steel and aluminum enclosure. All controls are located on the front panel along with two LEDs, one that indicates the power is on and the other flashes each time an output pulse is issued. This provides easy verification of correct operation.
Not For Human Use
701C Specifications
| |
Output Pulse Electrical Characteristics |
| Output Voltage |
0 to ±80V |
| Output Current |
0.4mA to 1A |
| Output Current Adjustment |
Front panel-mounted Current Range Switch and Adjustment potentiometer |
| Output Voltage Adjustment |
Front panel-mounted Voltage Range Switch and Adjustment potentiometer. |
| Pulse Width |
<50 µs to DC (controlled by External Trigger) |
| Pulse Rise and Fall Time |
<10 µs, 10-90% |
| Pulse Recurrence Frequency |
Single Shot to 200 kHz continuous |
| Maximum Duty Cycle |
20% |
| Maximum Average Power |
4 W |
| Pulse Output Connector |
BNC, front panel |
| |
Trigger |
| External Trigger Input |
+2V to +15V (unit protected from damage for input signals between -50 and +50 volts) |
| Trigger Modes |
Follow and Manual Trigger |
| Monitor Outputs |
| Sync Monitor |
+5V pulse generated for each output pulse. Monitor pulse length identical to output pulse length |
| Current Monitor (Optional) |
1 V per 1 A of output current |
| Voltage Monitor (Optional) |
1 V per 10 V of output voltage |
| |
General |
Dimensions (Excluding Connectors) |
9.5" W (half-rack mount) x 7" D x 3.4" (2U) H (24cm W x 17.8cm D x 8.6cm H) |
| Weight (approx.) |
6.4 lb (3.0 kg) |
| Input Power |
120 +/- 10% 50/60Hz, 1 amps max. 100,220 and 240VAC available |
| Fuse Type |
IEC 127-III. 5x20mm |
| Rating |
250VAC - T1.0A |
Click on the manual title to download it.
Instruction Manual - 701C Stimulator